Z-0714-2017: Peripherally Inserted Central Catheters (PICC), UPN H965750141, Catalog # 75-014 The Maximal Barrier Nurs

Recall record Z-0714-2017 for Angiodynamics Inc. (Navilyst Medical Inc.). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0714-2017
Company in FDA record
Angiodynamics Inc. (Navilyst Medical Inc.)
FDA record creation date
Dec 7, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.

Source facts

Action
Angiodynamics, Inc. (Navilyst Medical, Inc.) sent an Urgent Voluntary Medical Device Recall letter dated June 8, 2016, with reply forms to all affected customers. Customers were advised to immediately remove any affected product from inventory, segregate and secure in a location for return. Immediately forward a copy of the recall notification to all sites to which they have distributed affected product. If affected product is located in your institution, please call Navilyst Medical Customer Service at 1-800-772-6446 8am-7pm, EST Monday - Friday to obtain a replacement or credit for your returned product. Promptly complete, sign and return the Reply Verification Tracking Form by email to recall@angiodynamics.com or by Fax to 1-800-782-1357. For questions regarding this recall call 518-792-4112.
Initiated
2016-06-08
Root cause
Mixed-up of materials/components
Terminated
2018-06-15
Distribution
Nationwide Distribution to CA, FL, GA, HI, LA, MI, MN, MO, NV, OH & TX
Recall number
Z-0714-2017
Recall status
Terminated
CDRH record ID
148877
FDA record created
2016-12-07
Recall event number
74948

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗