Skip to content

Public FDA record

Z-0714-2024: Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

Recall record Z-0714-2024 for Ambu Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0714-2024
Company in FDA record
Ambu Inc.
FDA classification date
Jan 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

Source facts

Action
On December 5, 2023, the firm began notifying their customers via Urgent Field Safety Notice (Correction) letters. Customers were provided with an insert of Instructions for Use to include with their Instructions for Use booklet that highlights the new warning. Should you have additional questions, please contact Sanjay Parikh (sap@ambu.com) or Tammy Feyerherm (tfey@ambu.com).
Initiated
2023-12-05
Root cause
Device Design
Distribution
Domestic distribution nationwide.
Recall number
Z-0714-2024
Recall status
Open, Classified
CDRH record ID
204948
FDA classified
2024-01-11
Recall event number
93599

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗