Skip to content

Public FDA record

Z-0715-2008: Coloplast Elefant Suction/Irrigation Tubing, Part Number: AST 307, Coloplast Corp., Minneapolis, MN 5541

Recall record Z-0715-2008 for Coloplast Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0715-2008
Company in FDA record
Coloplast Corp
FDA classification date
Apr 3, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Lack of sterilization assurance: Sterilization of Elefant Suction-Irrigation Devices is not performed according to ISO 11135 standard.

Source facts

Action
Coloplast sent customers an "Urgent: Recall Notification" letter on 11/12/07. The letter described that the sterilization of the products was not performed according to ISO 11135 standard. The notification requested the customers to return the product immediately and to send a fax or contact the company via email in regard to inventory at hand. A phone script for distributors also included an advice to discontinue dispensing the product and to fax or email in regard to inventory at hand.
Initiated
2007-11-12
Root cause
Process design
Terminated
2008-11-04
Distribution
Nationwide, including Puerto Rico.
Recall number
Z-0715-2008
Recall status
Terminated
CDRH record ID
66572
FDA classified
2008-04-03
Recall event number
45907

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗