Skip to content

Public FDA record

Z-0715-2018: AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-ray

Recall record Z-0715-2018 for Mobius Imaging, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0715-2018
Company in FDA record
Mobius Imaging, LLC
FDA record creation date
Feb 24, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.

Source facts

Action
Mobius sent an Urgent Field Safety Notice letter dated September 26, 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Mobius Imaging is advising all customers to discontinue using the AIRO until update is completed on their system. Please provide a copy of this Field Safety Notification to all AIRO operators that use the AIRO system, and quarantine the system per hospital policies and procedures.
Initiated
2017-09-26
Root cause
Under Investigation by firm
Terminated
2018-08-16
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, CO, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, NC, NY, OH, OR, PA, RI, TN, TX, UT, VA, WA, WV and the countries of Argentina, Australia, Austria, Belgium, Canada, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Lithuania, Netherlands, Pakistan, Russia, Saudi Arabia, Spain, Switzerland, UAE, Uzebekistan
Recall number
Z-0715-2018
Recall status
Terminated
CDRH record ID
161461
FDA record created
2018-02-24
Recall event number
79159

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.