Skip to content

Public FDA record

Z-0715-2020: Arrow Berman Angiographic Catheter 50cc Item Number: AI-07130

Recall record Z-0715-2020 for Arrow International Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0715-2020
Company in FDA record
Arrow International Inc
FDA record creation date
Dec 19, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

Source facts

Action
Arrow International issued letters on 11/15/19 stating reason for recall, health risk, and action to take: immediately discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1- 855-419-8507, Attn: Customer Service or ema il recalls@teleflex.com. your local sales representative or Customer Service at 1-866-396-2111. .
Initiated
2019-11-15
Root cause
Process control
Terminated
2020-06-19
Distribution
U.S. Nationwide, Canada
Recall number
Z-0715-2020
Recall status
Terminated
CDRH record ID
178236
FDA record created
2019-12-19
Recall event number
84399

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗