Skip to content

Public FDA record

Z-0715-2024: (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)

Recall record Z-0715-2024 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0715-2024
Company in FDA record
Philips North America
FDA classification date
Jan 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)

Source facts

Action
Starting in July 2021, Philips Field Service Engineers (FSE) visited impacted customer sites to inspect the affected systems and turn on the circuit breaker of the extended battery of the UPS where required. All impacted systems were addressed as of February 9, 2023. Instructions for manufacturing and installation were updated on October 31, 2020, which included more detailed instructions for the correct setting of the circuit breaker.
Initiated
2021-07-15
Root cause
Process change control
Distribution
Domestic distribution to the following states and territories: CA DE FL GA IL KY MA ME MI MO NH NJ NM NY OH PA SC TN TX VA VT WI WY, and PR. International distribution worldwide.
Recall number
Z-0715-2024
Recall status
Open, Classified
CDRH record ID
205345
FDA classified
2024-01-11
Recall event number
93739

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗