- Recall classification
- Class II
Reason stated in the source
The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)
Source facts
- Action
- Starting in July 2021, Philips Field Service Engineers (FSE) visited impacted customer sites to inspect the affected systems and turn on the circuit breaker of the extended battery of the UPS where required. All impacted systems were addressed as of February 9, 2023. Instructions for manufacturing and installation were updated on October 31, 2020, which included more detailed instructions for the correct setting of the circuit breaker.
- Initiated
- 2021-07-15
- Root cause
- Process change control