Vapotherm humidification cartridges were found to contain microbial (Ralstonia spp.) cultures. This device uses cartridges from the same supplier.
Source facts
Action
Respironics contacted customers via tracked letter informing them to discontinue use of their NeoPAP system. Distributors were contacted via telephone and e-mail instructing them to contact their customers and instruct them to discontinue use of the NeoPAP system. The recalling firm issued a Customer Advisory Notice (CAN) dated 1/5/06 to all customers.
Initiated
2005-12-27
Root cause
Other
Terminated
2007-03-22
Distribution
The product was shipped to medical facilities in GA, IA, MA, MN, NJ, OK, PA, TX, and WI. The product was also shipped to Sweden, South Africia, Egypt, Greece, and Saudi Arbia.
Recall number
Z-0717-06
Recall status
Terminated
CDRH record ID
43762
FDA classified
2006-04-07
Recall event number
34390