Unnecessary operator exposure; X-ray machine may allow unwanted x-ray exposure to the operator.
Source facts
Action
On or about 9/6/2007, by Customer Notification Letter, users were notified that all affected units will be corrected by GE personnel with the addition of an additional shield.
Initiated
2007-09-28
Root cause
Other
Terminated
2011-11-01
Distribution
Worldwide- USA including states of CO, DE, FL, NJ, MO, NY, PA, and TX (with VA and Military facilities) and countries of Australia, Austria, Belgium, Brazil, Chile, China, The Czech Republic, Denmark, France, Germany, Hungary, Israel, Italy, Norway, Spain, Sweden, Switzerland, Thailand, and The UK.
Recall number
Z-0717-2008
Recall status
Terminated
CDRH record ID
67222
FDA classified
2008-03-15
Recall event number
46349