Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of
the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can
topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
Source facts
Action
Philips sent an Urgent Field Safety Notice letter December 20, 2017 dated to all affected customers. The letter identified the affected product, problem and the actions to be taken. For question contact your local Philips representative.
Initiated
2017-12-20
Root cause
Device Design
Terminated
2020-06-16
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-0717-2018
Recall status
Terminated
CDRH record ID
161321
FDA record created
2018-02-24
Recall event number
79091