Open original FDA record ↗Recall record
- FDA identifier
- Z-0718-2020
- Company in FDA record
- Bard Peripheral Vascular Inc
- FDA classification date
- Dec 19, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain product code/lot number combinations may experience introducer sheath compatibility issues when used with a 6F introducer sheath. Using the affected balloon expandable vascular covered stents with a 6F introducer sheath could result in difficulty or the inability to insert or advance the catheter through the introducer sheath. This could potentially lead to a slight prolongation of the procedure.
Source facts
- Action
- On 02/18/19, Urgent Product Defect Correction Notices were mailed to customers informing them to upsize to a 7F introducer sheath and to add relabeling stickers over the 6F introducer sheath size indication on the label. In addition, customers were asked to complete and return the Response Form. In addition, if the affected product has been further distributed, immediate action was requested to contact that location, advise them of the field correction notice and to forward the Urgent Product Defect Correction Notice to the affected customer. Customers having additional questions are encouraged to call 1-800-257-232.
- Initiated
- 2019-02-18
- Root cause
- Nonconforming Material/Component
- Terminated
- 2021-04-28
- Distribution
- 127 units were distributed in Australia to 32 Consignees.
- Recall number
- Z-0718-2020
- Recall status
- Terminated
- CDRH record ID
- 177990
- FDA classified
- 2019-12-19
- Recall event number
- 84173