Skip to content

Public FDA record

Z-0719-2018: MEDISTIM VERIQ SYSTEM, MODEL VQ4122 An intraoperative diagnostic system that utilizes ultrasonography to

Recall record Z-0719-2018 for Medistim Asa. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0719-2018
Company in FDA record
Medistim Asa
FDA record creation date
Feb 26, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A system component overheated causing melting of the connector and emission of smoke through the systems loudspeaker opening.

Source facts

Action
A letter was sent describing the device, reason for the recall, and risk involved. The firm stated that it would ship a VeriQ replacement system without the affected connector being used.
Initiated
2017-04-25
Root cause
Nonconforming Material/Component
Terminated
2024-05-03
Distribution
Distributed to New York.
Recall number
Z-0719-2018
Recall status
Terminated
CDRH record ID
160991
FDA record created
2018-02-26
Recall event number
78973

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗