Open original FDA record ↗Recall record
- FDA identifier
- Z-0720-2008
- Company in FDA record
- Varian Medical Systems Inc
- FDA classification date
- Feb 15, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect Software Validation- The ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery that will not be recognized properly by the system. Under these circumstances, an arc plan can be administered to a patient whereby the gantry will not rotate, resulting in delivery to a single location rather than through the gantry arc.
Source facts
- Action
- An Urgent Medical Device Correction letter was distributed October 2, 2007, to all direct consignees, informing them that the ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery will not be recognized properly by the System. The firm provided corrective instructions and informed consignees that a product modification will be developed to resolve the issue. A Varian Service Representative will be scheduling a visit to upgrade the application.
- Initiated
- 2007-10-02
- Root cause
- Software design
- Terminated
- 2011-05-17
- Distribution
- Worldwide Distribution
- Recall number
- Z-0720-2008
- Recall status
- Terminated
- CDRH record ID
- 65734
- FDA classified
- 2008-02-15
- Recall event number
- 45527