Z-0720-2021: MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibil

Recall record Z-0720-2021 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0720-2021
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Dec 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.

Source facts

Action
During the week of 12/01/2020, the Recalling Firm is sending an "URGENT MEDICAL DEVICE RECALL" Letter via e-mail and/or mail to customers informing them that due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification. Customer are instructed to: 1. Discontinue use and discard, per their laboratory procedure/guidelines, any remaining inventory of the affected lot. 2. The laboratory should consider reviewing and confirming organism identification and antimicrobial susceptibility test results of stains yielding results suggestive of a daptomycin "non-susceptible" category. 3. They should retain their inventory of other B1017-216 panel lots as they are not impacted by this issue. 4. Share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. If they have forwarded any of the affected product to any additional laboratories, they are instructed to provide the information/notification. 5. To assure receipt of the notification/important communication, customers are instructed to respond within 10 days in one of the following ways: -Electronically, if they received the communication via email -Manually complete and return the enclosed Response Form. The Recalling Firm is investigating the root cause of the issues and appropriate actions to prevent recurrence will be implemented. For questions, contact Customer Support Center at: -websited http://www.beckmancoulter.com -Phone 1-800-677-7226 in the United States -For customers in the United States, if replacement product is needed: --Complete the attached "Replacement Order Form" and email to askbeckman@ Beckman.com or fax to 866-294-7850 --Call Client Servic
Initiated
2020-12-02
Root cause
Process control
Terminated
2023-01-03
Distribution
U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, IN, KY, MA, MI, MO, NC, NH, NY, OH, OK, PA, SC, SD, TN, TX, VA, and WI. The countries of Australia, Austria, Bahamas, Bangladesh, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam.
Recall number
Z-0720-2021
Recall status
Terminated
CDRH record ID
184612
FDA record created
2020-12-23
Recall event number
86817

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗