Open original FDA record ↗Recall record
- FDA identifier
- Z-0721-2017
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA record creation date
- Dec 8, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.
Source facts
- Action
- Ray Search sent an Urgent Field Safety Notice Medical Device Correction letter (dated 11/8/2016) via e-mailed to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to Inspect their product and identify all installed units with the above software version number, then confirm you have read and understood this notice (contact information included).
- Initiated
- 2016-11-09
- Root cause
- Software design
- Terminated
- 2019-06-11
- Distribution
- US Nationwide Distribution in the states of California, Florida, Illinois, Louisiana, Michigan, New Jersey, Tennessee, Texas, Washington
- Recall number
- Z-0721-2017
- Recall status
- Terminated
- CDRH record ID
- 151111
- FDA record created
- 2016-12-08
- Recall event number
- 75731