There are lower than expected MICs for some gram negative species.
Source facts
Action
A customer letter was issued to impacted consignees on or around 01/10/2022. Customers are asked to check for affected product and the recalling firm requests that customers do not report results of cefiderocol when affected product is used with Thermo Scientific Mueller-Hinton Broth with TES, catalong T3462, lots 267261 to 402775.
Consignees are asked to disseminate the recall information within their organization and to any other organization to which affected products have been transferred.
An acknowledgement form should be submitted by email to MBD.vigilance@thermofisher.com.
Initiated
2022-01-10
Root cause
Under Investigation by firm
Terminated
2026-07-24
Distribution
Distribution throughout the United States.
Recall number
Z-0722-2022
Recall status
Terminated
CDRH record ID
191507
FDA record created
2022-03-04
Recall event number
89416