Z-0722-2024: MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Recall record Z-0722-2024 for Cook Medical Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0722-2024
Company in FDA record
Cook Medical Llc
FDA classification date
Jan 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Source facts

Action
The firm notified affected customers on December 7, 2023 via URGENT Medical Device Correction letters. Per the letter, customers can continue to use their MINC+ device whilst awaiting the defect correction, but Cook advises customers to be vigilant and monitor for device alerts. If the device produces audible and visual alerts to indicate an error with temperature, customers must immediately move any dishes to another incubator. If no other incubators are available, the device can be reset to normal operation by switching off mains power to the device for ten seconds then turning the power back on. The software/firmware update will undergo verification testing including compliance testing prior to release. The updated software/firmware is expected to be available on 01 February 2024. An authorized service agent will contact customers to arrange for the device to be corrected at their facility. The service agent will be updating the software/firmware of the device.
Initiated
2023-12-07
Root cause
Device Design
Distribution
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
Recall number
Z-0722-2024
Recall status
Open, Classified
CDRH record ID
205087
FDA classified
2024-01-12
Recall event number
93673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗