Open original FDA record ↗Recall record
- FDA identifier
- Z-0722-2026
- Company in FDA record
- DJO, LLC
- FDA classification date
- Nov 26, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.
Source facts
- Action
- Per recall notice, customers were asked to do the following: Users should immediately discontinue all further use of this product and discard any opened/used units. If you have unopened, unused units in your possession, please contact firm Customer Service at (844) 331-7767. You will be provided with a return authorization number and shipping label that is required for proper return and identification of the device. All users must complete the acknowledgement and response form and return to productsafety@enovis.com This notice must be distributed to all those within your organization and to any other organization, consignee, or user where these products were transferred. If you have any questions you can reach the firm via email at productsafety@enovis.com
- Initiated
- 2025-09-25
- Root cause
- Labeling Change Control
- Distribution
- US Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK.
- Recall number
- Z-0722-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216105
- FDA classified
- 2025-11-26
- Recall event number
- 97610