Open original FDA record ↗Recall record
- FDA identifier
- Z-0723-04
- Company in FDA record
- Abbott Laboratories HPD/ADD/GPRD
- FDA classification date
- Jul 20, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The package inserts contain incorrect control values for the urinary free cortisol extraction procedure.
Source facts
- Action
- Field Correction recall by Important Product Information letter dated 2/16/04, sent via e-mail to the Abbott international subsidiaries on the same date for notification of all of their accounts receiving the Cortisol reagents and X Systems Cortisol controls since March 2002. The accounts were informed of the incorrect values for the Cortisol controls, when using the urinary free cortisol extraction procedure, listed in the package inserts of the reagent kits as well as the controls. The letters contained a table listing the incorrect values and the correct values for each of the products, and the users were instructed to use the corrected values as provided in the letter.
- Initiated
- 2004-02-16
- Root cause
- Other
- Terminated
- 2004-07-26
- Distribution
- There was no domestic distribution. The product was distributed internationally through Abbott subsidiaries in Canada, United Kingdom, Japan, Singapore, Hong Kong, Korea, Australia, New Zealand, Honduras, Costa Rica, Panama, Dominican Republic, Germany and Pakistan.
- Recall number
- Z-0723-04
- Recall status
- Terminated
- CDRH record ID
- 31623
- FDA classified
- 2004-07-20
- Recall event number
- 28342