Skip to content

Public FDA record

Z-0723-04: TDx/TDxFLx Cortisol Reagent Pack, list 9116-65; Abbott Laboratories, Diagnostics Division, Abbott Park,

Recall record Z-0723-04 for Abbott Laboratories HPD/ADD/GPRD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0723-04
Company in FDA record
Abbott Laboratories HPD/ADD/GPRD
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The package inserts contain incorrect control values for the urinary free cortisol extraction procedure.

Source facts

Action
Field Correction recall by Important Product Information letter dated 2/16/04, sent via e-mail to the Abbott international subsidiaries on the same date for notification of all of their accounts receiving the Cortisol reagents and X Systems Cortisol controls since March 2002. The accounts were informed of the incorrect values for the Cortisol controls, when using the urinary free cortisol extraction procedure, listed in the package inserts of the reagent kits as well as the controls. The letters contained a table listing the incorrect values and the correct values for each of the products, and the users were instructed to use the corrected values as provided in the letter.
Initiated
2004-02-16
Root cause
Other
Terminated
2004-07-26
Distribution
There was no domestic distribution. The product was distributed internationally through Abbott subsidiaries in Canada, United Kingdom, Japan, Singapore, Hong Kong, Korea, Australia, New Zealand, Honduras, Costa Rica, Panama, Dominican Republic, Germany and Pakistan.
Recall number
Z-0723-04
Recall status
Terminated
CDRH record ID
31623
FDA classified
2004-07-20
Recall event number
28342

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗