Skip to content

Public FDA record

Z-0723-2022: CIRRUS HD-OCT, Models 5000 AngioPlex and 6000 AngioPlex

Recall record Z-0723-2022 for Carl Zeiss Meditec, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0723-2022
Company in FDA record
Carl Zeiss Meditec, Inc.
FDA record creation date
Mar 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Optic nerve head angiography scan to be turned off due to its distribution without pre-market clearance.

Source facts

Action
On 01/14/22, recall notices were mailed to customers who were asked to do the following: 1. Until further notice, please discontinue the acquisition of the ONH Angiography 4.5x4.5mm Scan. 2. Using the provided software kit(s), install software version 11.5.3 on all affected instruments. You will be receiving one software kit per affected instrument according to our records. An excerpt of the IFU has been included Section 4.7 Update Instrument Software, these instructions apply to all models. Refer to the IFU Section 4 for the full installation instructions. 3. Complete and return the Acknowledgement Form per the instructions in the attachment. In case you no longer have the device, we ask that you inform the recalling firm and forward this letter to the new owner. Customers with additional questions were encouraged to contact customer service at 1-800-341-6968 (select option 1, 1).
Initiated
2022-01-14
Root cause
No Marketing Application
Distribution
U.S Nationwide Distribution.: OH, CA, IN, AR, IA, LA, WI, OK, IL, OR, AL, KY, WY, WA, MN, NE, PA, SD, AZ, PR, TN, GA, FL, DE, NY, NH, TX, MS, MA, NV, UT, WV, ME, CO, AK, NM, KS, MO, MI, MD, ID, CT, NJ, MT, NC, VA, HI, VT, ND, SC, DC, RI
Recall number
Z-0723-2022
Recall status
Open, Classified
CDRH record ID
191855
FDA record created
2022-03-04
Recall event number
89524

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗