Skip to content

Public FDA record

Z-0723-2026: IntelePACS - InteleConnect / TechPortal

Recall record Z-0723-2026 for INTELERAD MEDICAL SYSTEMS INCORPORATED. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0723-2026
Company in FDA record
INTELERAD MEDICAL SYSTEMS INCORPORATED
FDA classification date
Nov 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.

Source facts

Action
On 10/17/2025, correction notices were emailed to customers who were asked to do the following: 1. Schedule Feature Release Upgrade (FRU) update immediately with the firm's Customer Success Manager (CSM). 2. Interim Mitigation: Until the update is applied, please advise all staff to verify the 'Relevant Clinical Info' field after any Case Editor modifications and consider temporarily documenting critical clinical information in alternative fields as backup. 3. complete and return the Acknowledgement Form via email to regulatory-affairs@intelerad.com. For more information or to ask any questions about the notification contact the firm's regulatory affairs department at regulatory-affairs@intelerad.com
Initiated
2025-10-17
Root cause
Software Design Change
Distribution
US Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.
Recall number
Z-0723-2026
Recall status
Open, Classified
CDRH record ID
216843
FDA classified
2025-11-26
Recall event number
97868

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗