Z-0724-2023: DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q

Recall record Z-0724-2023 for DeVilbiss Healthcare LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0724-2023
Company in FDA record
DeVilbiss Healthcare LLC
FDA classification date
Dec 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.

Source facts

Action
On November 18, 2022, the firm notified customers through Urgent Notice Medical Device Correction letters/emails. The firm is updating its instruction manual to include the following: " warnings related to applied parts that may come into contact with a patient that, when under single fault failure conditions while operating at maximum environmental temperatures, could cause the surface temperature of the part to exceed 41C; " clarifications regarding the expected service life for the unit; and " additional instructions regarding wear components that may require maintenance during the expected service life of the product. Customers may navigate to www.recallrtr.com/525series to obtain copies of the updated instruction manual. The firm may be reached at 1 (833) 408-0512, from 8:00 am to 7:00 pm EST Monday through Friday.
Initiated
2022-11-18
Root cause
Labeling False and Misleading
Distribution
US Nationwide. Foreign distribution pending.
Recall number
Z-0724-2023
Recall status
Open, Classified
CDRH record ID
196759
FDA classified
2022-12-19
Recall event number
91112

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗