Z-0724-2024: Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.

Recall record Z-0724-2024 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0724-2024
Company in FDA record
Maquet Cardiovascular, LLC
FDA classification date
Jan 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.

Source facts

Action
An Urgent Medical Device Correction notice dated December 1, 2023 was sent via FedEx 2-Day delivery. Please examine your inventory immediately to determine if you have any of the product codes and lot numbers listed in the notice and remove these from use. Return any unused/unexpired affected product to Maquet/Getinge. Please forward this information to all current and potential VasoView HemoPro EVH System users within your hospital / facility. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. If you have any questions, please contact your Maquet/Getinge representative or call the Maquet/Getinge Customer Support at 888-880-2874, Monday through Friday, between the hours of 6 AM to 5 PM (Pacific Standard Time). Whether or not your facility has affected product listed in the notice, please complete and sign the MEDICAL DEVICE REMOVAL - RESPONSE FORM to acknowledge that you have received the notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to VVHemoProParticulate2023.act@getinge.com or by faxing the form to 1-800-878-9105.
Initiated
2023-12-01
Root cause
Component design/selection
Distribution
US: OUS: China, Brazil, Hong Kong.
Recall number
Z-0724-2024
Recall status
Open, Classified
CDRH record ID
204971
FDA classified
2024-01-12
Recall event number
93615

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗