Skip to content

Public FDA record

Z-0724-2026: Xpert MTB/RIF. Model Number: GXMTB/RIF-US-10.

Recall record Z-0724-2026 for Cepheid. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0724-2026
Company in FDA record
Cepheid
FDA classification date
Nov 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product testing did not meet expected stability criteria.

Source facts

Action
Cephid notified consignees on about 11/05/2025 via FedEx letter. Consignees were instructed to ensure no affected units remain in inventory or are available for use since the affected lots have passed their expiration date. Past results obtained from Xpert MTB/RIF batch 1000812229 lot 41302 may be inaccurate. If these results were used to make clinical management decisions, it is recommended that the results be reviewed in the context of corresponding culture data, drug susceptibility testing, clinical presentation and other relevant laboratory information to ensure accurate patient care. If there is uncertainty or insufficient information to confirm the result, repeat testing to ensure accuracy. Consignees were requested to notify relevant clinical teams as needed. Cepheid will provide products for the purposes of retesting. Kits are expected to ship within 5 business days following receipt of the completed response form. Additionally, consignees were instructed to complete and return the Customer Response Form.
Initiated
2025-11-05
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution and Puerto Rico.
Recall number
Z-0724-2026
Recall status
Open, Classified
CDRH record ID
217135
FDA classified
2025-11-26
Recall event number
97911

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗