Z-0726-2022: HeartWare Ventricular Assist Device (HVAD) System

Recall record Z-0726-2022 for Heartware, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0726-2022
Company in FDA record
Heartware, Inc.
FDA record creation date
Mar 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.

Source facts

Action
Starting 27-Jan-2022 and ending 02-FEB-2022, Medtronic initiated an Urgent Medical Device Notice to global customers to inform them of the intent to update the cleaning instructions for power source connector pins on the HVAD" System Monitor AC Adapter, DC Adapter, and Batteries. The communication provided customers the updated instructions, the complaints received as well as the potential hazards of cleaning the power source connector pins. In addition to the Urgent Medical Device Notice a patient template was provided to facilitate the VAD Coordinators in notifying their patients. In the United States the notice was delivered to VAD Coordinators via UPS 2-day delivery. Consignees will be asked to confirm receipt of the letter and that they have acknowledge the actions taken for those nonimplanted, affected product. In addition, a courtesy copy was sent to shared care clinics for those affected patients. Outside the United States the notice will be delivered to consignees through a locally approved method. Consignees will be asked to confirm receipt of the letter and that they have taken the recommended action. Confirmation of the communication will be received by locally approved methods.
Initiated
2022-01-27
Root cause
Labeling design
Distribution
Distribution throughout the United States. OUS distribution in Argentina, Armenia, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea (South), Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritius, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Recall number
Z-0726-2022
Recall status
Open, Classified
CDRH record ID
191942
FDA record created
2022-03-04
Recall event number
89559

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗