Open original FDA record ↗Recall record
- FDA identifier
- Z-0726-2024
- Company in FDA record
- Limacorporate S.p.A
- FDA classification date
- Jan 12, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
Source facts
- Action
- On 8/29/23, the firm phoned all agents and surgeons to immediately stop surgeries with affected devices. On 10/10/23 the firm emailed recall notices to customers asking them to take the following actions: 1) Surgeons are suggested to perform a short-term patient s monitoring (radiographs at 6 weeks, 3-6 months and 1 year might be considered) and acting accordingly, based on a case-by-case evaluation. 2) Locate, quarantine, and return affected devices. Complete and return the response form to uscomplaint@limacorporate.com 3) Customers with questions can email the firm at medicalcomplaints@limacorporate.com
- Initiated
- 2023-08-29
- Root cause
- Device Design
- Distribution
- US: IL, MO, CA, IN, MI, OK
- Recall number
- Z-0726-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205077
- FDA classified
- 2024-01-12
- Recall event number
- 93668