Z-0726-2026: Aurora EV-ICD and Clinical EV-ICD, single chamber, extravascular implantable cardioverter defibrillators

Recall record Z-0726-2026 for Medtronic, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0726-2026
Company in FDA record
Medtronic, Inc.
FDA classification date
Nov 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated October 2025 was sent to customers. PATIENT MANAGEMENT RECOMMENDATIONS FOR PREVIOUSLY DISTRIBUTED DEVICES: Medtronic acknowledges that each patient requires unique clinical considerations. Based on internal investigation and consultation with our Independent Physician Quality Panel, Medtronic provides the following guidance: "Prophylactic device replacement is NOT recommended. "Schedule an in-clinic SmartSync interrogation to receive the software update, with the intent for all patients to receive theupdate within the next 3-6 months. For patients remotely followed via CareLink, your Medtronic representative can provide a report to assist with identifying patients that require this in-person update. You may contact your local representative to obtain an updated copy of the report at any time. CUSTOMER ACTIONS: "Ensure your SmartSync application is updated to include the Aurora EV-ICD Application. Contact your Medtronic representative if assistance is needed. "Please complete the enclosed Confirmation Form, following review of this letter, and email toRS.FCACRM@medtronic.com to confirm receipt of this notification. "Please forward this notice to everyone within your organization who needs to be aware and to locations where thesedevices could be followed if they are outside of your organization. If you have questions regarding this communication, please contact your Medtronic Representative.
Initiated
2025-10-29
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-0726-2026
Recall status
Open, Classified
CDRH record ID
217134
FDA classified
2025-11-28
Recall event number
97910

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗