Open original FDA record ↗Recall record
- FDA identifier
- Z-0727-03
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Apr 16, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
ICON 9.5 software does not apply flood corrections during whole body SPECT studies with multiple bed positions.
Source facts
- Action
- Siemens sent Customer Advisory letters dated 2/10/03 to all ICON 9.5 customers with Whole Body SPECT functionality on integrated camera systems including Diacam, Multispect 2, Multispect 3 and e.cam camera systems aquired on an ICON workstation. The accounts were informed of the problem related to applying the flood correction matrix to Whole Body SPECT studies in ICON version 9.5 software, and were provided the follwoing workaround: To ensure that the flood correction is applied, the user should acquire multiple single bed positions with the General SPECT predefine, indexing the bed manually between each SPECT acquisition. The data can then be combined into a Whole Body SPECT in the Whole Body SPECT Display program for review. In addition, users were requested to routinely perform standard quality control procedures at the facility to ensure proper image quality. No software upgrade is planned.
- Initiated
- 2003-02-10
- Root cause
- Other
- Terminated
- 2005-01-19
- Distribution
- Nationwide, United Arab Emirates, Argentina, Austria, Australia, Barbados, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Equador, Egypt, France, Germany, Great Britain, Georgia, Greece, Hong Kong, Israel, India, Italy, Jordan, Japan, Korea, Latvia, Mexico, Niger, Norway, New Zealand, Philippines, Pakistan, Poland, Russia, Saudi Arabia, SIngapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine and the Netherlands.
- Recall number
- Z-0727-03
- Recall status
- Terminated
- CDRH record ID
- 26005
- FDA classified
- 2003-04-16
- Recall event number
- 25541