General Electric Medical Systems Information Technology
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Non-invasive blood pressure readings on central station may not match bedside monitor due to software anomaly.
Source facts
Action
Consignees were notified via certified mail, return receipt requested, on Friday, February 27, 2004. GE Field service representatives will upgrade affected PatientNet Central Stations (RVS, IRVS, or Sitelink IRVS) to software version 2.01.
Initiated
2004-02-27
Root cause
Other
Terminated
2005-11-17
Distribution
Nationwide throughout the United States
Recall number
Z-0727-04
Recall status
Terminated
CDRH record ID
31796
FDA classified
2004-07-20
Recall event number
28410
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.