Mechanism failure: Medtronic has identified an issue with four lots of the Cardioblate Gemini-s Surgical Ablation Devices in which the polycarbonate distal coil retainer can fail during the course of a procedure, potentially resulting in an inability to apply the necessary jaw closure force.
Source facts
Action
Certified letters were sent on 11/21/07 to Consignees. The letter identifies the issue with the device and requests that customers return the device back to the firm. Customers will be requested to complete a Field Correction Action certificate acknowledging the receipt of the information. For additional information you can contact a Medtronic Sales Representative at 763-391-9167.
Initiated
2007-11-21
Root cause
Nonconforming Material/Component
Terminated
2011-12-17
Distribution
Nationwide Distribution including states of FL, IL, IN, LA, MI, PA, TN, TX, UT, and WI
Recall number
Z-0727-2008
Recall status
Terminated
CDRH record ID
66543
FDA classified
2008-02-02
Recall event number
45940