Reason stated in the source
When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
Source facts
- Action
- The firm disseminated the notices by mail on 11/12/2019. The letter identified the security vulnerability of certain GE Central Stations and ApexPro Telemetry servers and provided additional best practices, ensuring the following: 1. MC and IX Networks are isolated; 2. MC and IX Router/Firewalls block incoming traffic, as applicable; 3. Restricted physical access to Central Stations, Telemetry Servers, MC network and IX network; 4. Default passwords are changed as applicable; and 5. Password management best practices are adhered to Ensuring the networks are properly configured and isolated protects against these potential concerns and mitigates the risk.
- Initiated
- 2019-11-12
- Root cause
- Software in the Use Environment
- Terminated
- 2026-03-02
- Distribution
- Worldwide distribution.
- Recall number
- Z-0727-2020
- Recall status
- Terminated
- CDRH record ID
- 178167
- FDA record created
- 2019-12-20
- Recall event number
- 84372