Z-0728-2018: JANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended

Recall record Z-0728-2018 for Perkinelmer Life Sciences, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0728-2018
Company in FDA record
Perkinelmer Life Sciences, Inc.
FDA record creation date
Feb 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.

Source facts

Action
On August 4 2017 an URGENT Medical Device Correction letter titled "Horizontal Motion Error" was issued to customers informing them of the error notice and to follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. Question or concerns can be directed to Technical Support at 800-762-4000
Initiated
2017-08-02
Root cause
Nonconforming Material/Component
Terminated
2018-07-31
Distribution
Nationwide
Recall number
Z-0728-2018
Recall status
Terminated
CDRH record ID
161156
FDA record created
2018-02-27
Recall event number
79015

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗