Z-0728-2022: 1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscop

Recall record Z-0728-2022 for Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0728-2022
Company in FDA record
Stryker Corporation
FDA classification date
Mar 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.

Source facts

Action
Medical Safety Notification was issued On January 28, 2022. This is Part 1 of 2. Customer Instructions on January 28,2022 state: Check inventory to see if you have a 1688 Pendulum Camera. NO PRODUCT RETURN IS NECESSARY. Return the business reply form by email to endorecall@stryker.com to confirm receipt of this notification. Maintain awareness of this communication internally until part 2 of 2 of this field action is conducted. If you have any questions or concerns, please email endorecall@stryker.com. These solutions are temporary while Stryker Endoscopy develops a software update to the 1688 Camera Control Unit to solve this issue. Part 2 of the field action will be a software update on all 1688 CCUs, which Stryker will install. Temporary Solutions: a) Preventing the Flip: 1. The only way to prevent the possibility of the image flip is to remove the USB A-to-B Cable from the CCU to Hub. However, this will cause the loss of device control through the CCU. b) Methods to Avoid/Correct Flip: 1. Do not select "End Case" on the Hub while a camera head is inserted. 2. If the flip does occur, unplugging and re-plugging the camera head will resolve the issue. However, this is only temporary, and this may occur again if "End Case" is selected Recall notices dated May 6, 2022, were distributed to all customers with 1688 Camera Control Units. Customers were asked to check their inventory and if affected product is found, call their Sales Representative or OnSite Specialist to upgrade CCU Software. Complete and return the customer acknowledgment form. Please use Temporary Solutions (section above) until software upgrade is complete. The software version can be found on the bottom right corner of the screen during boot up.
Initiated
2022-01-14
Root cause
Software in the Use Environment
Distribution
Worldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL, IN, FR, ES, HK, IT, GB, CO, PE, MY, TH, CA, JP, MX, AR, NP, SG, PH, ZA, CL, KR, CN
Recall number
Z-0728-2022
Recall status
Open, Classified
CDRH record ID
191822
FDA classified
2022-03-07
Recall event number
89515

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗