During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Source facts
Action
Icotec issued a Field Action Notice to its consignee on 11/03/2022 by email. The notice explained the issue and that XXXX. The firm followed with an email to inform the consignee of their intent to recall the device. The notice requested the return of the device from the field. The firm will be replacing the devices.
Initiated
2022-11-03
Root cause
Nonconforming Material/Component
Terminated
2026-09-10
Distribution
US Nationwide distribution in the state of GA.
Recall number
Z-0728-2023
Recall status
Terminated
CDRH record ID
197182
FDA classified
2022-12-19
Recall event number
91174