Open original FDA record ↗Recall record
- FDA identifier
- Z-0729-2017
- Company in FDA record
- Merge Healthcare, Inc.
- FDA record creation date
- Dec 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the incorrect accession number (but still associated to the correct patient).
Source facts
- Action
- MERE sent an Urgent Medical Device Recall letter dated July 11, 2016, to all affected customers via email or via certified mail if they did not have an email address for the customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to respond to the notification using the enclosed form and the return addressed envelope. Customers were also instructed to ensure that all users of the product are provided with the notification and if product was further distributed customers need to identify their customers and notify them at once of the product recall. Customers with questions should send an email to recall@merge.com. For questions regarding this recall call 877-741-5369.
- Initiated
- 2016-03-28
- Root cause
- Software design
- Terminated
- 2020-03-20
- Distribution
- Nationwide Distribution
- Recall number
- Z-0729-2017
- Recall status
- Terminated
- CDRH record ID
- 151424
- FDA record created
- 2016-12-09
- Recall event number
- 75825