Z-0729-2022: HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the

Recall record Z-0729-2022 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0729-2022
Company in FDA record
Maquet Cardiovascular, LLC
FDA record creation date
Mar 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 mm L: 60 cm). Mislabeled product could cause a disruption of the vascular anastomosis in the long term, creating a pseudo-aneurysm

Source facts

Action
Getinge issued Urgent Medical Device Letter onJanuary 14, 2022, (via FedEx overnight) with adult signature required. Letter states reason for recall, health risk and action to take: Actions to be taken: " If you have already implanted the above-mentioned device, kindly notify us as soon as possible by contacting your Getinge representative or Customer Service (at 1-888-943-8872, options 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. You will be contacted shortly by our Medical Affairs Officer in order to discuss the appropriate follow up. " Please examine your inventory immediately to determine if you have the HEMASHIELD GOLD Product Knitted Microvel Double Velour Straight Vascular Graft (D: 7 mm; L: 60 cm) SN 1318567653, Ref Number M002020952070. " Should you have the un-used affected product, you are eligible for credit upon receipt of Response Form (see page 4). " Please contact Getinge Customer Service at 1-888-943-8872, (press option 2) Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone), to request shipping instructions to return any affected product. " Pack the product to be returned with the appropriate return documents. " Whether you have affected product or not, please complete and sign the attached URGENT MEDICAL DEVICE REMOVAL RESPONSE FORM (page 4) to acknowledge this recall and return the completed form to Getinge by e-mailing a scanned copy to Hemashieldgold- Recall.us@getinge.com or by FAX to 8772811346. " Please forward this information to all current and potential Hemashield Gold Knitted Microvel Double Velour Vascular Graft users within your hospital / facility. " If you are a distributor who has shipped any affected product to customers, please forward this document to their attention for appropriate action. If you have any questions, please contact your Getinge representative or, for technical questions, please contact Customer Service (at 1-888-943-8872, options 2), Monday th
Initiated
2022-01-14
Root cause
Packaging process control
Terminated
2023-09-19
Distribution
US Nationwide distribution in the state of MO.
Recall number
Z-0729-2022
Recall status
Terminated
CDRH record ID
191965
FDA record created
2022-03-07
Recall event number
89573

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗