Open original FDA record ↗Recall record
- FDA identifier
- Z-0730-2022
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Apr 5, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.
Source facts
- Action
- Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers. The letter inform customers that the firm will distribute a software correction via update instruction at no cost to the customers.
- Initiated
- 2022-03-07
- Root cause
- Software design
- Distribution
- US Nationwide, Worldwide distribution
- Recall number
- Z-0730-2022
- Recall status
- Open, Classified
- CDRH record ID
- 192328
- FDA record created
- 2022-04-05
- Recall event number
- 89738