Open original FDA record ↗Recall record
- FDA identifier
- Z-0731-2017
- Company in FDA record
- Elekta, Inc.
- FDA record creation date
- Dec 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Edits to Particle field definition parameters may not be saved when the field definition window is saved.
Source facts
- Action
- The firm, Elekta, sent an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter dated 11/2016 was sent to affected customers on 12/2/2016. The letter describes the product, problem and actions to be taken. The customers were instructed to save and re-open the field definition form when manual edits have been performed to confirm that the edits were save; post this notice in a place accessible to all users; advise the appropriate personnel working with the product the content of the letter; and complete and return the Acknowledgement Form via fax to: 408 830 8003 immediately upon receipt but no later than 30 days. Elekta will resolve the problem with the MOSAIQ 2.60 SP7F, MOSAIQ 2.60 SP14 and 2.64 SP3. Customers are encouraged to contact Elekta support to get the corrected version. If you have any questions, contact RA Manager Americas at 1-770-300-9725 or email to: Elekta_Global_PMS@elekta.com.
- Initiated
- 2016-12-02
- Root cause
- Software design
- Terminated
- 2021-11-05
- Distribution
- Worldwide Distribution-US (nationwide) to states of: FL, IL, MO, NJ, OH, OK, TN, TX, VA, and countries of: Italy, Japan, South Korea, Sweden and United Kingdom.
- Recall number
- Z-0731-2017
- Recall status
- Terminated
- CDRH record ID
- 151455
- FDA record created
- 2016-12-09
- Recall event number
- 75853