The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.
Source facts
Action
The customer was contacted via email on 10/16/18.
Initiated
2018-10-16
Root cause
Under Investigation by firm
Terminated
2020-06-29
Distribution
The products were distributed to the following US states: TN.
Recall number
Z-0731-2019
Recall status
Terminated
CDRH record ID
169625
FDA record created
2019-01-12
Recall event number
81723