Z-0731-2024: (1) Patient Information Center iX, Model No. 866389 (2) Patient Information Center iX Expand, Model No. 8

Recall record Z-0731-2024 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0731-2024
Company in FDA record
Philips North America
FDA classification date
Jan 13, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.

Source facts

Action
On November 28, 2023, the firm notified affected consignees through URGENT Medical Device Correction letters. This failure is applicable to Apple iOS devices only. If customers use iOS devices (e.g. iPhone, iPad, etc.) to receive event notifications: 1) Please conduct patient monitoring at the central station or patient bedside. 2) Please do not use iOS event notifications for patient monitoring. A Philips representative will contact you to schedule a visit to install a software patch to your device which will correct the current issue. Patch PIC iX 4.2.2 will be provided for PIC iX 4.X and patch CareEvent C.03.07 will be provided for CareEvent C.03.X. If you need any further information, please contact your local Philips representative, or call 1-800-722- 9377.
Initiated
2023-11-28
Root cause
Software design
Distribution
Worldwide distribution: US (Nationwide) to states of: CA, GA, IA, MA, MI, MN, NJ, NM, NY, OH, PA, TN, TX, WA, WV; and OUS (Foreign) distribution to countries of: Australia, Austria, Belarus, Belgium, Brazil, Canada, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, India, Indonesia, Ireland, Italy, Japan, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Poland, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Utd.Arab.Emir., and Uzbekistan
Recall number
Z-0731-2024
Recall status
Open, Classified
CDRH record ID
204999
FDA classified
2024-01-13
Recall event number
93638

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗