Inconsistent data - A buffer overflow may occur between information system and picture system if the system experiences a high load. (e.g. RIS/PACS communication is unusally high)
Source facts
Action
The recalling firm issued a Customer Safety Advisory, via certified mail. to the affected customers on 11/27/07. The letter informs the customer of the issue and also provides instruction to detect and avoid its occurrence. In addition, the firm is developing a software update to correct the issue, which will be installed on the affected systems when it is released.
Initiated
2007-11-27
Root cause
Software design
Terminated
2008-09-17
Distribution
Nationwide Distribution, including states of AL, MO, NC, NY, OH, OR, TN, and TX.
Recall number
Z-0732-2008
Recall status
Terminated
CDRH record ID
67179
FDA classified
2008-02-23
Recall event number
46318