Open original FDA record ↗Recall record
- FDA identifier
- Z-0732-2019
- Company in FDA record
- Meridian Bioscience Inc
- FDA record creation date
- Jan 14, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An enzyme reagent included in the kit is not maintaining stability through claimed product expiration, resulting in the potential for false negative results when tested with patient specimens and the positive control.
Source facts
- Action
- Urgent: Voluntary Medical Device Recall notification letters dated 12/17/18 were distributed to customers. The letter identified the affected product, problem and actions to be taken. The letters instructed customers to review inventory for affected lots, forward the letter to any customers, contact the firm replacement product, complete the response form. For questions contact Meridian's Technical Service Department by telephone at 800-343-3858 or 513-271-3700 or by facsimile 513-272-5432.
- Initiated
- 2018-12-17
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2020-04-30
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AK, CA, CO, DE, IL, KS, KY, LA, MD, MO, NC, ND, NJ, NY, SC, TX, and VA. The products were distributed to the following foreign countries: India and Italy.
- Recall number
- Z-0732-2019
- Recall status
- Terminated
- CDRH record ID
- 169846
- FDA record created
- 2019-01-14
- Recall event number
- 81787