Skip to content

Public FDA record

Z-0732-2025: 20/30 Priority Pack Accessory Kit w/Copilot, REF: 1003327, used for cardiovascular procedures.

Recall record Z-0732-2025 for Abbott Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0732-2025
Company in FDA record
Abbott Vascular Inc
FDA classification date
Dec 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.

Source facts

Action
On 11/6/2024, recall notices were distributed to customers asking them to take the following actions: 1) Stop using affected devices. 2) Complete and return the acknowledgement form. 3) Return unused devices to the firm. 4) Share this notification with relevant personnel in your organization. 5) If you have further distributed/transferred the affected products, notify those customers. If you have any questions contact firm customer service at 800-227-9902
Initiated
2024-11-06
Root cause
Component change control
Distribution
US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.
Recall number
Z-0732-2025
Recall status
Open, Classified
CDRH record ID
211209
FDA classified
2024-12-20
Recall event number
95707

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗