Skip to content

Public FDA record

Z-0733-2019: LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm

Recall record Z-0733-2019 for Waldemar Link GmbH & Co. KG (Mfg Site). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0733-2019
Company in FDA record
Waldemar Link GmbH & Co. KG (Mfg Site)
FDA record creation date
Jan 14, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.

Source facts

Action
Urgent Medical Device Recall notification letters dated 12/7/18 were distributed to customers. Please identify your total inventory for this product. Please take the following actions: 1) Immediately cease use and segregate the above described product. 2) Complete the enclosed Consignee Reply Form, even if you have no inventory. 3) Return the above product and a copy of the Consignee Reply Form using Federal Express Account #: 007183879 to: LinkBio Corporation 101 Roundhill Drive Rockaway, NJ 07446 ATT: Nikhil Mangale Recall Return Authorization #: R-2018-03
Initiated
2018-12-07
Root cause
Under Investigation by firm
Terminated
2019-12-17
Distribution
US Nationwide in the states of: FL, GA, IA, IN, KS, MI, MN, and TN.
Recall number
Z-0733-2019
Recall status
Terminated
CDRH record ID
169803
FDA record created
2019-01-14
Recall event number
81778

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗