Skip to content

Public FDA record

Z-0734-2014: Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs

Recall record Z-0734-2014 for Remel Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0734-2014
Company in FDA record
Remel Inc
FDA classification date
Jan 14, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes.

Source facts

Action
Oxoid and Remel part of Thermo Fisher Scientific sent an Urgent Medical Device Recall letter dated November 20, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The letter asked customers to destroy the affected product, if any was found on hand. The letter also asked if the customer further distributed the product that they inform their downstream customers. A response form was included with the letter and the letter instructed customers to complete and return the response form to indicate they had received the notification and that they had complied with it. If you have any questions please contact our Technical Service Department at 800-255-6730 (US) or 913-888-0939 (International).
Initiated
2013-11-21
Root cause
Unknown/Undetermined by firm
Terminated
2014-12-11
Distribution
Nationwide Distribution including the states of AL, AZ, CA, CO, FL, HI, IA, IL, KY, LA, MD, MI, MN, MO, MS, NC, NY, OH, SC, TN, TX, VA, WI, and WV
Recall number
Z-0734-2014
Recall status
Terminated
CDRH record ID
124706
FDA classified
2014-01-14
Recall event number
67099

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.