Open original FDA record ↗Recall record
- FDA identifier
- Z-0734-2017
- Company in FDA record
- Accuray Incorporated
- FDA classification date
- Dec 12, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Accuray has become aware of a potential safety issue involving unexpected treatment robot motion when removing an accessory. This issue is limited to certain CyberKnife Systems with software version 1 0.6.
Source facts
- Action
- Accuray sent an Urgent Safety Advisory Notification letter dated November 22, 2016, to all affected customers. Letters alerted users to the issue and the sequence of events necessary for it to occur. Accuray reps will visit all sites to perform a software upgrade to correct the issue. For more information or any questions about this notification, please contact Accuray Customer Support at 1-877-668-8667 or visit the Accuray Technical Solution Center at http://www.accuray.com/services-support. For further questions, please call (408) 716-4600.
- Initiated
- 2016-11-22
- Root cause
- Software design
- Terminated
- 2017-05-22
- Distribution
- Worldwide Distribution - US Distribution to the states of : PA, WA, MT, NY, SC, CA, MN, AK, NJ, FL, KY, WI., and to the countries of : Bulgaria, Canada, France, Germany, India, Japan, Luxembourg, Netherlands, Portugal, Switzerland and Turkey
- Recall number
- Z-0734-2017
- Recall status
- Terminated
- CDRH record ID
- 151477
- FDA classified
- 2016-12-12
- Recall event number
- 75863