Skip to content

Public FDA record

Z-0734-2023: 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy

Recall record Z-0734-2023 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0734-2023
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Dec 20, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Source facts

Action
Fresenius issued Updated Urgent Medical Device Correction REV B Customer Notification letter (Field Action Number: FA-2020-09-W _REV B) to consignees that have CDX equipped machines on 07-27-2022 via USPS First Class Mail. Letter states reason for recall, health risk and action to take: First, a software issue in SW versions 2.72 and earlier. All machines should be updated to current software (SW version 2.74), pursuant to the July 20, 2021, letter. Second, hardware issue that affects only CDX machines. To reduce the chance of EMI causing a false Remove USB Device 2 alarm from occurring we suggest the following: " Insert a USB device in USB Port 2 that will continually draw current. For example, a blank USB memory stick or a USB powered card reader. o NOTE: DO NOT use passive USB devices in USB Port 2 that go into a suspended state drawing minimal/low current which can trip the sensing circuit. An example would be a USB corded optical mouse. Complete/Return the response form enclosed.
Initiated
2022-07-27
Root cause
Process design
Distribution
US Nationwide distribution.
Recall number
Z-0734-2023
Recall status
Open, Classified
CDRH record ID
197048
FDA classified
2022-12-20
Recall event number
91154

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗