Open original FDA record ↗Recall record
- FDA identifier
- Z-0735-2022
- Company in FDA record
- Ortho-Clinical Diagnostics, Inc.
- FDA record creation date
- Mar 8, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.
Source facts
- Action
- On December 10, 2021, the firm notified consignees through ORTHO PLUSSM (a service that provides customers with Ortho Clinical Diagnostics product notifications electronically) of the product issue. Customers were instructed to discontinue use and discard the affected lot. Customers should forward the notification if the product was distributed outside of their facility.
- Initiated
- 2021-12-10
- Root cause
- Nonconforming Material/Component
- Distribution
- Domestic distribution nationwide. Foreign distribution to Canada.
- Recall number
- Z-0735-2022
- Recall status
- Open, Classified
- CDRH record ID
- 191333
- FDA record created
- 2022-03-08
- Recall event number
- 89362