Z-0736-2012: Varian brand Eclipse Treatment Planning System, 7.3, 8.0, 8.1, 8.2, 8.6, 8.9, 10.0; Model Number: H48; R

Recall record Z-0736-2012 for Varian Medical Systems, Inc. Oncology Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0736-2012
Company in FDA record
Varian Medical Systems, Inc. Oncology Systems
FDA classification date
Jan 13, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An anomaly has been identified with the Eclipse Treatment Planning System when planning for the Siemens Multileaf Collimator (MLC) where a virtual jaw can be positioned inside the MLC aperture and then used as the beam limiting device for the dose calculation. This will result in an incorrect dose distribution within Eclipse.

Source facts

Action
Varian Medical Systems sent a Urgent Medical Device Correction letter dated December 12, 2011, to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to pay careful attention to any warnings displayed regarding the Virtual Collimator Jaws when the plan is calculated. If the warning indicates that the Virtual Collimator Jaws are inside the MLC aperture, do not proceed with the calculation. Varian is continuing to evaluate possible technical solutions for this issue. Special Instructions for customers outside the USA and Canada: In order to satisfy regulatory requirements, we request that you complete the attached Proof of Notification or Receipt Verification Card once you have read this document and return it to Varian Medical Systems. "We sincerely apologize for any inconvenience and thank you in advance for your cooperation." Please call for USA and Canada: 1.888.VARIAN5 (888.827.4265) Europe: +41 41 749 8844.
Initiated
2011-12-12
Root cause
Software design
Terminated
2014-08-15
Distribution
Class II Recall - Worldwide Distribution -- USA (nationwide) including the countries of Canada and Europe.
Recall number
Z-0736-2012
Recall status
Terminated
CDRH record ID
106612
FDA classified
2012-01-13
Recall event number
60837

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗