Skip to content

Public FDA record

Z-0736-2024: PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill),

Recall record Z-0736-2024 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0736-2024
Company in FDA record
Becton Dickinson & Company
FDA classification date
Jan 17, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.

Source facts

Action
An URGENT: Medical Device Product Advisory dated 11/16/23 was sent to customers. Please Take the Following Actions: 1. Immediately review your inventory for the affected product. a. If the product is used for syringe pumps, destroy all affected product subject to this field action following your institution's process for destruction. b. If the product is used for manual application, the product can continue to be used per IFU. 2. If the affected product was previously used on a patient, no further actions are needed. 3. If the affected product is currently in use with a syringe pump, replace the syringe with a non-impacted lot. Please note that healthcare facilities and providers should rely on their syringe pump manufacturer's IFU for the specific syringes that are compatible with their syringe pumps. 4. Share this field action notification with all users of the product within your facility network to ensure they are also aware of this field action. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement if product is used on syringe pumps and has been discarded. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA 1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787 Distributors Please Take the Following Actions: 1. Immediately review your inventory for the affected product. Destroy all unused product subject to the product advisory following your institution s process for destruction. 2. Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide
Initiated
2023-11-16
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.
Recall number
Z-0736-2024
Recall status
Open, Classified
CDRH record ID
204809
FDA classified
2024-01-17
Recall event number
93562

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗