Skip to content

Public FDA record

Z-0737-05: FlexTrayTM Endopath Gastric Tray, Kit code KDG05.

Recall record Z-0737-05 for Ethicon Endo-Surgery. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0737-05
Company in FDA record
Ethicon Endo-Surgery
FDA classification date
Apr 27, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.

Source facts

Action
The recalling firm sent letters to hospitals and distributors, dated 12/14/04.
Initiated
2004-12-15
Root cause
Other
Terminated
2012-05-30
Distribution
The devices were distributed to hospitals and distributors located throughout the United States and to foreign consignees located in Saudi Arabia, Austria, Egypt, United Kingdom, Germany, France, Lebanon, The Netherlands, Spain, Pakistan, South Africa, Italy, United Arab Emirate, Czech Republic, Sweden, Switzerland, Greece, India, Portugal, Slovenia, Hungary, Palestine, and Kuwait.
Recall number
Z-0737-05
Recall status
Terminated
CDRH record ID
36832
FDA classified
2005-04-27
Recall event number
30727

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗